| MRI of the Brain | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
3.2
Clinical Indications for MRI of the Brain The
common clinical indications for brain MRI include: · Multiple Sclerosis (MS) · Primary tumour assessment and /or metastatic disease · AIDS (toxoplasmolisis) · Infarction (cerebral vascular accident (CVA) versus transient ischemic attack (TIA) · Haemorrhage · Hearing loss · Visual disturbances · Infection · Trauma ·
Unexplained neurological symptoms or deficit.
(Westbrook, 1999) 3.3
Image interpretation of brain MRI The following is a summary of MRI Tissue characteristics of brain structures adapted from Woodward (2001). Table
1 Tissue characteristics on MRI Relaxation time Image Contrast
_______________________
____________________________
Being
given the normal appearance of brain structures on MRI we can now move a
step further and try to identify any abnormalities in the underlying
images. This may be
performed by the process of patterned recognition.
Woodward (2001) presents three rules of thumb, which may help
with the search of abnormal areas: ·
Are the right and
left halves of the brain symmetrical? Both
halves of the brain should appear symmetrical, if not there is a good
chance it’s an abnormality ·
Are there areas
that exhibit any signs of mass effect? Pathological
processes as tumours, usually take more space than the original
structure, either by destroying it or pushing it aside.
The ventricles should be intact and symmetrical and the sulci and
gyri must be clear, without swelling and distortion. ·
When looking at
T2-weighted images are there any bright signal intensities other than
the CSF spaces and venous structures? Most
pathological processes involve the presence of oedema, which appears
bright on T2-weighted images. The
table below will help in the recognition of normal anatomy and pathology
based on MRI contrast characteristics devised Westbrook (2001). Table
2 Pathology characteristics on MRI
3.4
Image optimisation in MRI of the brain 3.4.1
Equipment For
the above indications a dedicated head coil (quadrature or phased array)
is essential in routine MRI if the brain.
Most head coils are now equipped with mirrors or prisms allowing
vision of the exterior. Pads
and straps are usually essential for patient immobilisation and earplugs
are supplied to prevent any damage to the ears from the gradient noises.
For centers performing Echo-Planar Imaging (EPI), diffusion and
perfusion imaging high performance gradients are required (Westbrook,
1999). 3.4.2
Image quality factors The
most common factors, which should be taken into consideration when
discussing MR image quality, include signal-to-noise (SNR) Raito, image
contrast and spatial resolution (Edelman, Hesselink and Zlatkin, 1996). The
relationship between image data and noise is expressed as SNR.
An image having a high SNR is usually sharp and appears less
grainy. It is therefore
recommended to,
(Edelman
et al, 1996 p 61) Many
factors have an effect on the SNR in MR imaging. A higher magnetic field strength will produce a higher SNR,
especially when motion artefacts are suppressed. Low bandwidth pulse sequences yield images with a high SNR
when compared to high bandwidth sequences. The uses of spin echo pulse
sequences increases the SNR when compared to gradient echo (GRE)
sequences, due to the 90 flip angles that are used were the resultant
echo has greater signal amplitude.
As the Number of signal averages (NEX) increases so does the SNR
however, this also increases the scan times. Coherent
noise (e.g. respiratory or pulsatile movements) and incoherent noise
(e.g. RF leak) both reduce the SNR.
Moreover, receiver coil sensitivity governed by its design and
receiver coil distance from the tissue of interest will also affect the
SNR resolution (Edelman et al, 1996; Mitchell, 1999; Westbrook, 1993). The
spatial resolution of an image determines the viewer’s ability two
points as separate and distinct and is determined by voxel size and SNR.
If the spatial resolution is increased, the SNR decreases in
direct proportion to the volume of the voxel.
Therefore, a coarse matrix, a large field of view (FOV) and a
selection of thick slices will improve the SNR but will unfortunately
yield a poor spatial resolution (Edelman
et al, 1996; Mitchell, 1999; Westbrook, 1993). Tissue
contrast refers to the differences in signal intensity from different
tissue structures. Tissue
contrast (e.g. T1 or T2 contrast) may be manipulated through the
selection of scan parameters and pulse sequences.
Some factors include the repetition time (TR), echo time (TE),
matrix, NEX, slice thickness, orientation and location and bandwidth.
The magnetic field strength and the coil also effect image
contrast (Edelman et al, 1996). 3.4.3
Routine brain MRI The
SNR and contrast characteristics in the brain are usually good.
A dedicated quadrature head coil is used yielding a high and
uniform signal, and good spatial resolution at relatively short scans
times. The routine brain
examination at St. Luke’s Hospital includes the following: ·
3-planar T2* FGRE ·
Axial PD & T2 FSE ·
Axial FLAIR irFSE ·
Coronal T1 SE ·
Sagittal T2 FSE A
standard study usually starts with a scout multislice acquisition using
a fast gradient echo sequence (FGRE), with T1-weighted contrast (Jager
et al, 2001). This
is followed a SE/FSE axial dual echo sequence providing proton density
contrast on the first echo and T2 contrast on the second echo. The first spin echo is generated early by selecting a short
TE. Since only little T2
decay has occurred, T2 contrast between tissues is insignificant in this
echo. The second spin echo
is generated much later by using a long TE, whereby T2 contrast is now
much greater. In dual echo
sequences the TR is long in order to minimise T1 contrast in different
tissues (Westbrook and Kaut, 1993). Diagram
3 A Spin Echo PD/T2 (dual echo pulse sequence)
SE/FSE
sequences are used since they produce relatively good grey/white matter
contrast at relatively short scan times.
Since the TR used is of the range 2000-2400ms, CSF signal
intensity increases due to its high proton density that may reduce the
contrast between CSF and periventricular lesions (e.g. MS).
This explains the use of the dual echo sequence (Westbrook,
1999). Many
centers add a sagittal SE/FSE T2-weighted acquisition in routine brain
MRI because they feel that this aids detection of corpus callosum
involvement in MS (Jager et al, 2001). For T2 weighted images a TR of 4000ms or higher may be used
with an echo train length (ETL) of 16 or more as opposed to an ETL of 4
in the dual echo acquisition. However,
as the ETL increases the possibility of image blurring also increases
(Westbrook, 1999). Westbrook
(1999) also suggests the use of a coronal SE/FSE PD/T2 (dual echo) as
part of routine brain imaging. Brown
and Semelka (1999) state that a number of innovations have been
developed for brain examinations. These
include magnetisation transfer suppression (ideal for background water
suppression), diffusion and perfusion weighted images (sensitive in
acute stroke) and echo train spin echo (ETSE) fat suppression imaging. In our MRI unit at St Luke’s Hospital, the FLAIR sequence
is used in routine brain imaging. FLAIR
is a variation of the inversion recovery pulse sequence. In flair, the signal from CSF is nulled, by selecting an
inversion time (TI) corresponding to the time of recovery of CSF from
the 180-degree pulse to the transverse plane and there is no
longitudinal magnetisation present in CSF.
A TI of 200ms is usually required to suppress the CSF signal. Since the CSF signal is nulled pathologies adjacent to the
CSF are seen more clearly e.g. MS (Westbrook
and Kaut, 1993). 3.4.3.1 Fast imaging techniques There
are two ways to speed scan times in MRI. Primarily, the amount of data collected can be reduced or
secondly, to reduce the time span to acquire the number of data points
needed. The former can be
achieved by using conjugate symmetry or a rectangular FOV where only a
portion of the raw data is sampled while maintaining resolution through
zero filling (Woodward, 2001). Data collection time can be reduced by a number of ways, for example by reducing the NEX
and the matrix size. However,
as stated in section 3.3.2 SNR and /or resolution may be affected. Reduction in scan time may be achieved by using fast spin
echo (FSE) sequences. Woodward
(2001) states that FSE uses a modification of the SE sequence, where
multiple data lines per sequence is acquired. Diagram4 The Fast Spin Echo Pulse sequence
RF=radiofrequency,
SS=slice select, PE=phase encoding, FE=frequency encoding
Technique: As in
routine spin-echo imaging, a 90-degree RF pulse is followed by a single
180 degree refocusing pulse. However, in fast spin echo (FSE) imaging,
multiple 180 degree refocusing pulses are performed with multiple
resulting spin-echoes. Each echo has a different phase encoding value
and is assigned to a different k-space line. As the number of echoes
acquired after a 90-degree excitation increases, scan time decreases but
signal to noise also decreases and T2 blurring increases. Maximum ETL is
limited by true T2 decay. Typical
TR:
>3000 ms (for T2-weighted images) TE:
Effective TE is time when the central k-space views are acquired (which
determines image contrast). ETL: Echo
Train Length - generally between 4 and 16
(MARS, 2002) As
in routine spin-echo imaging, a 90-degree RF pulse is followed by a
single 180 degree refocusing pulse. However, in fast spin echo (FSE)
imaging, multiple 180-degree refocusing pulses are performed with
multiple resulting spin-echoes. Each echo has a different phase encoding
value and is assigned to a different k-space line. As the number of
echoes acquired after a 90-degree excitation increases, scan time
decreases but signal to noise also decreases and T2 blurring increases.
Maximum ETL is limited by true T2 decay (General Electric, 2002). Moreover, Woodward (2001) states that the successive
180-degree echoes in FSE reduce its sensitivity to haemorrhage. To compensate for this factor, modifications involving the
incorporation of GRE in addition to the SE have been proposed to
increase the level of magnetic susceptibility effects. 3.4.4
Advanced Fast Imaging techniques 3.4.4.1
Echo planar imaging (EPI) Diagram
5 Single and Multi Shot Echo Planar Imaging RF=radio
frequency, SS=slice select, PE=phase encoding, FE=frequency encoding
Technique: Instead of
acquiring spin echoes by using consecutive 180 degree refocusing pulses
(as in FSE imaging), the sign of the readout (frequency-encoding)
gradient is rapidly alternated and gradient echoes are acquired,
resulting in markedly reduced scan times. MS-EPI requires several
excitations to cover the 64 to 128 views of k-space, while "single
shot" EPI does this with one excitation, but requires faster and
stronger gradients. Spin-echo EPI images can also be acquired, by
rapidly switching gradients in a similar manner, this time following the
180 degree refocusing pulse. Typical
MS-EPI: #
of shots required is k-space views desired / ETL GRE
readout: ETL 4, TE (effective) 5 ms, echo spacing 5 ms SE
readout: ETL 4, TE (effective) 80 ms, echo spacing 5 ms Typical
Parameters: MS-EPI:
# of shots required is k-space views desired / ETL GRE
readout: ETL 4, TE (effective) 5 ms, echo spacing 5 ms SE readout: ETL 4, TE (effective) 80 ms, echo spacing 5 ms (MARS,
2002) Below
are some advantages and disadvantages of EPI adapted from Mitchell,
1990):
3.4.5
Functional imaging techniques – Diffusion and Perfusion imaging The
random motion of molecules in fluids is responsible for molecular
diffusion. These motions are
however independent from the rotational motions responsible for T1
recovery and T2 decay. Contrast
from molecular diffusion is caused because phase coherence of in vivo
water protons undergoing random motion is reduced resulting in MR signal
attenuation. By the
application of strong encoding gradients in particular directions, signal
attenuation can be exploited to produce valuable diffusion variations.
In order to accomplish diffusion weighed images, the attenuation
factor B is calculated and used to produce the diffusion based MR signals.
B = exp (-bD) Where
b is a factor depending on the gradient intensity and duration and D is
the diffusion coefficient (Woodward, 2001).
Therefore, by increasing the b value the amplitude of the diffusion
gradient will also increase making the acquisition more sensitive to the
diffusion of water in normal brain. The
signal from areas with normal diffusion is low while areas of restricted
diffusion appear bright (Westbrook, 1999). “In
this way diffusion weighted MRI (DWI) can be used for the early detection
of acute strokes, ischemic lesions and other brain and body
abnormalities.” (Woodward, 2001 p137) DWI
is also very sensitive in the early detection of oedema, ischaemia, cancer
and cyst differentiation. Unfortunately DWI are very sensitive to motion and therefore
more prone to artefact production. Hence,
the shortest sequences possible as the Single-shot EPI is recommended at
the detriment of the in-plane resolution and SNR (Woodward, 2001). Westbrook
(1999) refers to perfusion as the microcirculation or the delivery of
blood to the tissues. Perfusion
imaging refers to the to the measurement of blood volume in these areas.
To measure perfusion either stationary spins need be suppressed or
signal intensity from perfusing spins enhanced.
Either employing motion sensitive gradients or injection of
contrast agents may achieve these (Westbrook and Kaut, 1993). In
order to maximise the perfusion effect rapid bolus delivery is essential
were an injection rate of 2-3mls/s is recommended.
Perfusion images are usually achieved through SE-EPI or GRE-EPI
sequences. The latter greatly increases sensitivity to susceptibility
effects of the bolus of gd-DTPA as it enters the brain on the first pass.
However, air/tissue interface susceptibility artefacts are much
more severe with the GRE-EPI than with the SE-EPI sequence.
The SE-EPI is suitable for brain perfusion imaging, providing
temporal resolution of enhancing lesions and indicating activity.
However, since SE-EPI is not so sensitive to Gd-DTPA susceptibility
a higher dose may be needed (Westbrook, 1999).
3.4.6
Artefacts in brain MRI Many
different forms of artefacts exist in MRI.
The main source of artefacts in brain imaging is from flow motion
of the carotids and vertebral arteries (Westbrook, 1999). Because blood flows throughout the body, different spins
constantly flow into and out of the imaging planes. Therefore, as the protons absorb the energy, they are not
available to return a signal back into the original location and a flow
phenomenon is created (Woodward, 2001).
This artefact may be reduced by sending a spatial pre-saturation
pulse is placed inferior to the FOV.
If the FOV is large an inferior pre-saturating pulse is sufficient,
as there is no flow coming in the FOV from any other direction. For small FOVs, superior, inferior, right and left spatial
pre-saturation pulses are necessary (Westbrook, 1999). Gradient
Moment Nulling (GMN), which
is a flow compensation technique, minimises artefacts especially in the
posterior fossa. However, it
not only increases the signal in the vessels but minimises the TE
available. Therefore, it is
reserved for T2 and T2* weighted sequences (Westbrook, 1999).
Another method of minimising the flow motion artefact is through
ECG or pulse triggering. Imaging
is synchronised with the cardiac cycle so that blood or CSF motion is
nearly identical from one phase encoding step to the next. Therefore, the ghosting that is generated in the along the
phase encoding axis is eliminated. The
major problem with this technique is that it is time-consuming (Bogdan,
1999). Another method is to
swap the phase encoding axis with the frequency-encoding axis in order to
remove the ghosting from the region of interest (Westbrook, 1999). Another
source of artefact in brain MRI is patient motion. Artefacts due to patient motion may be due voluntary and
involuntary. Voluntary motion
can be usually reduced by making the patient feel more comfortable and
restriction the motion through immobilisation.
Involuntary motion occurs due to respiration, cardiac motion,
orbital motion and swallowing. The
latter may be reduced by gating and pre-saturation techniques (Bogdan,
1999). Westbrook (1999)
recommends the use of fast sequences for the uncooperative patient.
However, FSE while faster, may produce more artefacts when compared
to SE because one of the central k-space lines is being filled during each
TR period. On the other hand,
single shot FSE (SS-FSE) is a good sequence for reducing motion artefacts,
where a whole brain scan can take up to 30s using ETL of up to 128. If some degree of patient motion is still present, SS-EPI may
be used which on the other hand is prone to air/tissue magnetic
susceptibility artefacts (Westbrook, 1999). Other
artefacts may be present in brain MRI imaging, which may be
sequence/protocol, related. On
of these are the aliasing artefacts.
The techniques used for spatial localisation assign a unique
frequency and phase too each location within the image.
The acquisition matrix and the desired FOV in the phase encoding
and readout gradients determine this.
This artefact occurs when the FOV is smaller than the anatomical
region and the frequencies for this tissue exceeds the Nyquiest limit,
causing frequency wraparound. This
effect may be eliminated by low-pass filters placed in the receivers
passing frequencies below the cut-off frequency and attenuating those
above. Aliasing artefacts can
also occur in the phase encoding direction when protons outside the FOV
are excited. When these
protons undergo phase changes corresponding to the frequency changes the
artefact occurs. Solutions
for this artefact may be by simply selecting a larger field of view or the
application of pre-saturation pulse to region outside the FOV (Bogdan,
2001; Brown and Semelka, 1999). The
chemical shift artefact may be also encountered in brain MRI.
This artefact appears as a displacement of the fat signal along the
frequency-encoding direction relative to other tissues.
This occurs because the protons in fatty acids have a resonant
frequency slightly different from the protons in water molecules.
Visually, the image shows a thin, bright band at one edge of a fat
–fluid interface and a dark band at the edge of the opposite side.
The chemical shift artefact may be reduced by either suppressing
the fat signal or using the STIR technique (Bogdan, 2001; Brown and
Semelka, 1999). Another
artefact, which may occur, is referred to as cross talk.
This occurs when there is contamination by an RF pulse in one
slice, which may affect the adjacent slice.
This can reduce the longitudinal magnetisation and thus the
decreased signal intensity the resulting image from the contaminated
slice. This effect can be
reduced by interleaving the slice acquisitions giving enough time for the
contaminated slice to return to the equilibrium value.
Cross-talk results in reduced SNR, particularly for tissues with
long T1 (e.g. CSF), since these recover more slowly from cross-excitation
(Mitchell, 1999). Substances
with weak magnetic properties (paramagnetic) cause magnetic susceptibility
artefacts, resulting from alterations of the local magnetic field within
the patient. This artefact is
most prominent in non-spin echo sequences although it may be visualised on
T2-weighted spin echo images as dark regions (low signal) at the foci of
the paramagnetic materials. On
non-SE sequences, e.g. GRE, these foci bloom and appear larger.
Also, these materials can cause marked distortion of the local
magnetic field, resulting in signal dropout and the distortion of adjacent
anatomy (Cardoza and Herfkens, 1996). 3.4.7
Contrast usage in Brain MRI Gd-DTPA
is a T1-shortening contrast agent currently in widespread use in MRI
departments. The following
list indicates those structures normally enhanced by Gd-DTPA: ·
Dural sinuses - superior
sagittal sinus transverse
sinuses ·
Nasopharynx mucosa ·
Choroid plexus ·
Pituitary gland ·
Infundibulum ·
Cavernous sinuses (Woodward,
2001). Contrast
studies are usually required for tumour assessment such as meningiomas and
neuromas. Active MS plaques
and metastases also enhance, especially after high doses. At St. Luke’s Hospital, certain radiologists prefer to use
a triple dose of Gd-DTPA as they believe that tumour enhancement is
improved. Infectious process,
such as abscesses, encephalomyelitis and toxoplasmosis are very
susceptible to enhancement. Contrast
studies will also help the early detection of these infections and improve
the patient outcome (Carolan and Runge, 1992).
In addition, the meninges also enhance so that infectious
tuberculosis, leptomeningeal tumour spread and post-trauma to the meninges
can be visualised. Contrast
is very useful in determining the age of infarcts.
Recent infarcts can enhance while maximum enhancement occurs when
the blood-brain barrier has been breached.
Old and chronic infarcts do not enhance
(Westbrook, 1999). SE or 3D incoherent (spoiled) GRE are the sequences of choice after contrast administration. In 3D SPGR the data may be reformatted in the desired plane or location after the scan is complete. Perfusion imaging, with a rapidly infused bolus of contrast is useful in measuring the activity level of a lesion (Westbrook, 1999). Following the administration of Gd-DTPA, the use of magnetisation transfer (MT) may increase the conspicuity of contrast enhancing lesions e.g. active MS plaques (Woodward, 2001).
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||